 |
 |
Inlet |
n/a |
Ionisation |
MALDI |
Analyzer |
TOF |
Detector |
EM |
Resolution |
12,000 |
Detection Limit
(ppt) |
n/a |
m/z Range |
500,000 |
m/z Range Detail |
n/a |
Sensitivity |
n/a |
Sensitivity Detail |
n/a |
Accelerating Voltage |
n/a |
Dynamic Range |
n/a
|
Features: |
-
Linear and Harmonic Reflectron operation
- SNP analysis
- Neutral Detection
- Collision Cell, Pulsed Extraction and Post Source Decay (PSD)
- Quadratic Field design enabling seamless PSD in a single scan
- 2D rotating sample stage for high throughput automated analysis
- Integrated sample viewing camera and laser optics to give maximum control of experiments
- Dual microchannel plate detector for reflectron operation
- Automatic calibration for maximum precision in peptide measurements
- Suitable for industrial polymer analysis
- Vacuum System: differential pumping by two turbomolecular pumps (70 l/s and 250 l/s) coupled to a two stage rotary pump (0.9 m3/h)
- All within a footprint of 63 x 58 x 150 cm (W x D x H)
- Data system hardware: network ready integrated pentium PC, 512M RAM, 80 Gb hard drive
- Data system software: full control of all MS parameters, automated sequence programming for batch operation, integrated camera and viewing software, report generator, proteomics software, rapid seamless de-novo sequencing and peptide/PSD DB
- Extra statistics:
Reflectron Resolution >12,000 (FWHM), ACTH(MH+2465) Mass Accuracy External calibration 100ppm RMS Mass Accuracy Internal calibration <10ppm RMS Precursor Ion Selection 100 (FWHM) Sensitivity < 1fmol
|
Options: |
|
|
|
|
|