Inlet |
n/a |
Ionisation |
MALDI |
Analyzer |
TOF |
Detector |
EM |
Resolution |
2,500 |
Detection Limit
(ppt) |
n/a |
m/z Range |
500,000 |
m/z Range Detail |
n/a |
Sensitivity |
n/a |
Sensitivity Detail |
n/a |
Accelerating Voltage |
n/a |
Dynamic Range |
n/a
|
Features: |
-
Linear MALDI-TOF with pulsed extraction
- Ideal for low cost analysis of peptides, proteins, oligonucleotides and synthetic polymers
- Vacuum System: turbomolecular pump (70 l/s) coupled to a two stage rotary pump (0.7 m3/h)
- All within a footprint of 63 x 58 x 150 cm (W x D x H)
- Data system hardware: network ready integrated pentium PC, 512M RAM, 80 Gb hard drive
- Data system software: full control of all MS parameters, automated sequence programming for batch operation, report generator, peptide DB, proteomics and synthetic polymer software
- Extra statistics:
Resolution >2,500 (FWHM), Substance P(MH+1347) Mass Accuracy External calibration 500ppm RMS Mass Accuracy Internal calibration 100ppm RMS Sensitivity < 1fmol
|
Options: |
-
2-D Sample Stage
- High Mass Detector
|
|
|
|
|