Home Products Demonstrations Links Contacts
 
1200L Triple Quad
Varian
Request Demo Get Quote
Inlet LC Ionisation APCI, ESI
Analyzer MSn, Quadrupole Detector HED
Resolution 1,500 Detection Limit (ppt) n/a
m/z Range 1,490 m/z Range Detail 10 - 1500 Da
Sensitivity 50 Sensitivity Detail 50:1 (ESI MSn: 10pg in 5µL Reserpine at 200µL/min)
Accelerating Voltage n/a Dynamic Range 10 5

Features:

  • Scan modes: Full scan, SIM, Neutral loss, SRM, MRM

  • Long 185mm path ensures high dissociation efficiency

  • Lensless design between quadrupoles simplifies tuning

  • Ion detector for + or - ions

  • Vacuum System: dual stage, N2 , air-cooled turbomolecular pumps 230 l/s and 170 l/s

  • Data System Software: Manual and automated quantitative analysis, fast reporting and network compatible

  • Extra Statistics:
    Scan rate: 500 - 6000 u/sec
    Resolution: unit mass resolution over entire range so Max Res = 1500/1 = 1500
    Mass axis stability: +- 0.1u over 24h

  • Applications: Pharmaceutical, clinical, toxicological, environmental

Options:

  • NIST/Wiley/PMW spectral DB

  • Autosampler

  • GC interface giving EI, CI, NCI