Inlet |
LC |
Ionisation |
APCI, ESI |
Analyzer |
MSn, Quadrupole |
Detector |
HED |
Resolution |
1,500 |
Detection Limit
(ppt) |
n/a |
m/z Range |
1,490 |
m/z Range Detail |
10 - 1500 Da |
Sensitivity |
50 |
Sensitivity Detail |
50:1 (ESI MSn: 10pg in 5µL Reserpine at 200µL/min) |
Accelerating Voltage |
n/a |
Dynamic Range |
10 5
|
Features: |
-
Scan modes: Full scan, SIM, Neutral loss, SRM, MRM
-
Long 185mm path ensures high dissociation efficiency
-
Lensless design between quadrupoles simplifies tuning
-
Ion detector for + or - ions
-
Vacuum System: dual stage, N2 , air-cooled turbomolecular pumps 230 l/s and 170 l/s
-
Data System Software: Manual and automated quantitative analysis, fast reporting and network compatible
-
Extra Statistics:
Scan rate: 500 - 6000 u/sec
Resolution: unit mass resolution over entire range so Max Res = 1500/1 = 1500
Mass axis stability: +- 0.1u over 24h
-
Applications: Pharmaceutical, clinical, toxicological, environmental
|
Options: |
-
NIST/Wiley/PMW spectral DB
-
Autosampler
-
GC interface giving EI, CI, NCI
|
|
|
|
|