Double-focussing arrangement: single 750-mm radius, 70-degree magnet with a 943-mm radius, 70-degree ESA
-
Sixteen Faraday detectors: fixed central 10-array and three each to the low and high mass side (recording high dispersion)
-
Ion counting system records and limits the use of detectors to protect them
-
Independent computer controlled adjustable slits positioned in front of all detectors, allow movement in width and central position providing perfect alignment of the peaks of interest
-
Fully adjustable source defining slit and selectable alpha slit allow maximum transmission for desired resolution
-
Resolves both low and high mass side interferences essential for laser ablation analysis of the lighter elements
- Vacuum System: 3 turbo pumps and three high vacuum ion pumps on analyzer
-
Vacuum System: 3 turbo pumps and three high vacuum ion pumps on analyzer
-
Data System Software: MS software for instrument tune-up, monitoring and analysis (inc. time resolved analysis)
-
Applications: Isotopic ratio determination
|